Most of Radiometer’s analysers have automatic QC (quality control). This reduces the risk of errors, saves valuable time and provides laboratory staff with a powerful set of tools to establish and verify accurate and precise analyser performance, securing compliance at all times.
ABL blood gas analysers
The ABL series of blood gas analysers automatically perform QC, check the analytical system, initiate necessary corrective actions and document all activities.
Everything from inlet to waste is controlled in this process, and assessed across QC lot changes via QC data logs, QC statistics and Levey-Jennings plots.
QC measurements are automatically evaluated against calibration status, analyser malfunctions and QC control ranges, and given a status to provide the operator with an overall analyser parameter status.
If a QC result is outside the control range, the analyser can repeat the QC measurement automatically while attempting to remedy the situation. The parameter can also be automatically locked out, requiring operator intervention.
AQT90 FLEX immunoassay analyser
The AQT90 FLEX analyser is programmed to automatically perform three types of checks. Each check is performed with a relevant frequency to establish and verify accurate and precise analyser performance.
The environmental check ensures that the analyser is ready for testing, and that consumables have not expired. If an environmental check fails you will not be able to start a test.
The process check makes certain that errors have not occurred during each and every testing process. If a process check fails you will not get a result.
The pre-analysis check ensures that only tests compatible with the chosen anticoagulant can be selected.